Welcome to the official website of BEIJING YUYAN BIOTECH CO., LTD!
Location> Solution > Blood Transfusion

Contact Us

BEIJING YUYAN BIOTECH CO., LTD

Tel:+86 13552014099 ( Wechat / Whatsapp )

E-mail:sales@yuyan-biotech.com
quanquan1965@163.com

Addr:J122,CHANGHONG WEST ROAD NO.73, XILU STREET, FANGSHAN DISTRICT, BEIJING,

Frozen Plasma Thawer
(Cryoprecipitate Thawer)
Model:BSLCD-I-28

  • Hits:68
  • Time:2025-9-4

Details

General Features
Used to thaw clinical plasma. It is suitable for thawing frozen plasma for cryoprecipitate coagulation factors preparation (centrifugation method).It can thaw fresh frozen plasma by constant temperature water bath, monitor and record the temperature changes in the whole process of plasma thawing in real time, and realize the data traceability of thawing process.The two key control points during cryoprecipitate preparation are water bath temperature and thawing time.
Vibration and Shake System
During the thawing process, the heat exchange efficiency between plasma and water was accelerated by shaking the system at a fixed frequency.
The plasma bag was slightly squeezed by the circulating water through the shaking system to shorten the thawing time.
Water Circulation System
Large water storage capacity.Low water temperature fluctuation.
While standby, the heating water tank internal water temperature equilibrium by the internal water circulation system.
During thawing, the thawing tank and heating tank equilibrate the temperature of the plasma thawing environment through the water circulation system.
Optimized refrigeration system
Adopts tubular heat exchange, which can effectively increase the heat transfer area and raise heat transfer efficiency.
Adopts micro-channel condenser, compact structure, high energy efficiency ratio.
Adopts small closed AC compressor,small volume, low noise.
Plasma Temperature Control after Thawing
After thawing, the water circulation between the thawing tank and the heating tank is non-stop to ensure that the temporary storage temperature of plasma is maintained at 1 -6 '℃, reducing the risk of loss of coagulation factors caused by long-term storage of plasma in room temperature.

Quality Control Function

The quality control function ensure the safety and effectiveness of the temperature control system.

Data Management
A simulated plasma bag can be intervened to effectively monitor the plasma core temperature.

A color LCD touch screen is used to record the thawing temperature and thawing time of plasma in real time.The data related to plasma thawing will be record automatically and can be accessed at any time.The scanner is optional to scan and automatically record the bar code of the thawed plasma bag, and add bar code records to the temperature/time curve to meet the needs of data traceability and quality management.

Technical Parameters

Max. amount of thawing

28 bags (200ml)

Thawing temperature range

1℃~6°C

Timing function

10min ~180min

Water pump capacity

60L/min

Power supply/KW

AC 220V 50Hz/ 4000VA

Weight

140Kg

Contraindication: None. Use according to the instructions and under the guidance of a physician.