Tel:+86 13552014099 ( Wechat / Whatsapp )
E-mail:sales@yuyan-biotech.com
quanquan1965@163.com
Addr:J122,CHANGHONG WEST ROAD NO.73, XILU STREET, FANGSHAN DISTRICT, BEIJING,
Automated Blood typing Analyzer is suitable for Micro-column gel method,which is simple to operate, high sensitivity, can detect low content of antigen and antibody,the results are clear and reliable, easy to store, and meet the requirements of a variety of data management.
Analyzer is suitable for automatic detection of forward and reverse ABO blood group typing, Rh blood group identification,irregular antibody screening and cross-matching in various hospitals, blood banks and centers of disease control and prevention.
Sample scanning,micro-column card loading,sample addition, dilution reagent addition, incubation, centrifugation and result interpretation were all completed automatically without manual intervention.
General Features
Supporting Micro-column Gel Card
Technical Parameters
|
Cardina-3000A |
|
|
Test Project |
1.ABO blood group positive typing test;2.ABO blood group reverse typing test;3.Rh(D)blood group typing test;4.Cross matching of blood test;5.Antibody Screening test;6.Rh typing test |
|
Functional Use |
Sample scanning, micro-column loading, sample addition, dilution, reagent addition, incubation, centrifugation, and result interpretation were all completed automatically without manual intervention. |
|
Mode of Operation |
Cyclic injection, card continuous injection; Multi-item parallel detection can customize the test item of each sample, and submit ABO forward and reverse typing, Rh D typing, irregular antibody, cross matching and other test items for a batch of samples at the same time. |
|
Test Carrier |
Micro-column gel card (use both 6-hole and 8-hole micro-column gel cards) |
|
Speed of Processing |
Blood typing test ≥600 per hour Antibody screening test ≥ 576 per hour |
|
Position of Sample |
96.The original specimen tube was loaded on the machine. Automatic entry and exit of the sample holder. Sample cycle replenishment |
|
Reagent Position |
14,Automatic mixing reagent.Automatic identification of reagent position |
|
Position of Card |
144. Can be loaded once |
|
Bar Code Scanning |
Sample and micro-column gel card and reagent bottle automatic scanning. Support a variety of bar codes |
|
Robotic Arm |
Three independent robotic arms, two are sampling arms, and one is robotic arm |
|
Sample Loading Channel |
Two sample loading channel, Adopts permanent sampling needle to avoid the cost of sampling needle consumables, and has the function of blocking needle alarm |
|
Mechanical Grasper |
One independent mechanical claw, with card drop alarm function |
|
Accuracy of Sampling |
Adding amount(μl) CV True degree 10 ≥3% ≤1μl 50 ≥2% ≤2μl |
|
Amount of Liquid Added |
1μl~ 500μl |
|
Position of Incubation |
Four independent incubation tables. 48 incubation sites total. Target temperatures ≤±2℃ |
|
Centrifugal Position |
Two 12 card independent centrifuges.Rotating speed 0-3800 RPM |
|
System of Interpretation |
High-resolution color CCD imaging interpretation, Real and clear pictures. The original image can be saved permanently |
|
Bar Code Printer |
Reagent name and validity printing |
|
Mode of Operation |
Support pre-centrifugation of micro-column card, Continuous supplementary loading of sample and reagent and micro-column card, Emergency insertion |
|
Puncture |
Independent puncture module, Puncture and sample addition and card removal were carried out simultaneously and side by side without interference |
|
Liquid Volume Detection |
The Buid volume was monitored in real time and displayed in the software interface Automatic alarm when liquid volume is insufficient or waste liquid volume is excessive |
|
Protection of Security |
Fully closed appearance structure, Dual alarm system with sound and indicator light |
|
Operate System and Software |
WIN11 Chinese operating system, MYSQL data base, can be connected with LIS/ HIS system to achieve two-way communication |
|
Operate Environment |
Temperature 15°C~ 30°C Humidity30%-75% |
|
Power |
AC 220V±22V 50Hz±1Hz |
|
Instrument Size |
920*760*1800mm (Length*Width*Height) |
|
Weight |
200kg |