Tel:+86 13552014099 ( Wechat / Whatsapp )
E-mail:sales@yuyan-biotech.com
quanquan1965@163.com
Addr:J122,CHANGHONG WEST ROAD NO.73, XILU STREET, FANGSHAN DISTRICT, BEIJING,
Features
Stable and reliable
The basket-type design ensures there are no complex moving parts inside the container. All motors are located in the normal temperature area, making them less prone to damage.
Multi-device networking
The side setting of the transfer container transportation channel can be connected by light track to build a multi-machine networking system, which can combine multiple automatic liquid ammonia containers into an integrated automatic warehouse.
Full-process cold chain
The entire process of storage, vial picking, and transportation is carried out at -150°C, which ensures that biological samples do not exceed the glass transition temperature.
Six-axis vial picking
The six-axis robotic arm is capable of picking up cryogenic vials of various brands and types. It features a torque feedback function, enabling real-time monitoring of tube drop anomalies.
Ultra-low consumption
The small-bore wide-mouth container design features a container opening size that is less than 20% of the container's cross-sectional area, resulting in a static liquid nitrogen consumption as low as 12 L per day.
|
Item |
BB66K-196 |
BB126K-196 |
Unit |
|
Dimension |
1320*1880*2460 |
1640*2100*2700 |
mm(W*D*H) |
|
Storage capacity (0.75 mL) |
65664 |
126720 |
kg/m2 |
|
Storage capacity (2 mL) |
25536 |
50688
|
pcs |
|
Storage temperature |
≤-185 |
≤-185 |
°C |
|
Vial-picking temperature |
≤-150 |
≤-150 |
°C |
|
Sample transport temperature |
≤-150 |
≤-150 |
°C |
|
Static duration |
15 |
20 |
d |